When not using the Parastep, store all
components in a safe place to prevent
damage. Electrodes
- Electrodes, under normal use, are
expected to provide 10 to 15
applications. The user should note the
date when beginning to use each
package of electrodes on the electrode
package. The personeal electrodes may
have a shorter expected life than the
shin, quadriceps, or hip electrodes
because of their positioning.
- Follow use and storage instructions
on the electrode package to extend the
life and improve the performance of
electrodes.
Electrode Leads and Power Cable
- Record the date of first use of
electrode leads on the label located
in the center junction of each lead.
- Visually inspect leads, cables,
locking mechanisms, and connecting
plugs before each session of use.
- Always plug electrode leads into
electrodes carefully, while grasping
both the electrode socket and the lead
pin. Make sure the connection is snug,
but do not use excessive force.
- If an electrode lead or the power
cable is suspected to be a source of
system failure, check its output with
the Paratester. If it tests defective,
immediately discard it and replace it
with a new lead or cable.
- If an electrode lead or power cable
plug connector begins to lose its
ability to lock securely into the jack
in the stimulator/control unit or the
battery pack, or if the lock release
pin breaks, immediately discontinue
using it and replace it with a new
lead or cable. This situation should
not occur as long as you protect the
lead and cable connectors with the
protective covers when you are not
using the system.
WARNING: DO NOT ATTEMPT TO MAKE
ELECTRODE LEAD OR POWER CABLE REPAIRS BY
YOURSELF AS SERIOUS PERSONAL INJURY COULD
RESULT.
Walker
- Monthly examine all joints of
walker, tighten as necessary, and
lubricate with 3-in-one household oil.
- Check rubber boots and wheels of
walker for wear. Replace as necessary.
For maintenance, service or technical
support requirements you may have, please
contact our Technical Support Department
at:
Technical Support
Email: fzeiss@ameritech.net
Phone: (937) 439-9131
Fax: (937) 439-9272 |